SSPC-PA 9
July 20, 2015
Editorial revisions August 8, 2015 and February 12, 2018
2
4. Gage Requirements
The ultrasonic gage shall be capable of measuring the
thickness of a coating over a non-metallic substrate. To qualify
under this standard, the gage shall have a minimum accuracy
within ±5 percent. All gages shall be calibrated and adjusted
following the manufacturer’s instructions and be in accordance
with ASTM D 6132.
5. Calibration and Verication of Accuracy
5.1 ASTM D6132 describes three operational steps
necessary to ensure accurate coating thickness measurement:
calibration, verication of accuracy and adjustment of coating
thickness measuring gages, as well as proper methods for
obtaining coating thickness measurements. These steps shall
be completed before taking coating thickness measurements to
determine conformance to a specied coating thickness range.
Verication of accuracy shall be performed in accordance with
ASTM D6132.
5.2 Gages shall be calibrated by the equipment
manufacturer, its authorized agent, or an accredited calibration
laboratory. A test certicate or other documentation showing
traceability to a national metrology institution is required.
There is no standard time interval for re-calibration, nor is
one absolutely required. Calibration intervals are usually
established based upon experience and the work environment,
or when specied. A one-year calibration interval is a typical
starting point suggested by gage manufacturers.
5.3 To guard against measuring with an inaccurate gage,
gage accuracy shall be veried according to the procedures
described in ASTM D6132. Verication shall be performed, at
minimum, at the beginning and end of each work shift. The
user is advised to verify gage accuracy during measurement
acquisition (e.g., hourly) when a large number of measurements
are being obtained. If the gage is dropped or suspected of
giving erroneous readings during the work shift, its accuracy
shall be rechecked.
5.4 Record the serial number of the gage, the reference
standard used, the stated thickness of the reference standard
as well as the measured thickness value obtained, and the
method used to verify gage accuracy. If the same gage,
reference standard, and method of verication are used
throughout a job, the information need be recorded only once.
The stated value of the standard and the measured value shall
be recorded each time accuracy is veried.
5.5 If the gage fails the post-measurement accuracy
verication check, all measurements acquired since the
last accuracy verication check are suspect. In the event of
physical damage, wear, or high usage, or after an established
calibration interval, the gage shall be rechecked for accuracy
of measurement. If the gage is not measuring accurately, it
shall not be used until it is repaired and/or recalibrated (usually
by the gage manufacturer).
6. Measurement Procedures
6.1 Verify that the gage will provide accurate readings of
the coating to be measured, using the procedure in Note 9.2.
6.2 NUMBER OF READINGS: Repeated gage readings,
even at points close together, often differ due to small surface
irregularities of the coating and the substrate (see Notes 9.4
and 9.5). Therefore, a minimum of three (3) gage readings
shall be made for each spot measurement of the coating. For
each new gage reading, move the probe to a new location
within the 15-cm (6-inch) diameter circle dening the spot.
Each gage reading shall be at least 5 cm (2 inches) from the
other gage readings. Discard any single unusually high or low
gage readings that are not repeated consistently. The average
of the acceptable gage readings is the spot measurement. See
Note 9.6 regarding removal of couplant after completing spot
measurements.
6.3 NUMBER OF MEASUREMENTS: Unless otherwise
specied in the procurement documents (project specication),
an area measurement is obtained by taking ve (5) separate
spot measurements (average of the gage readings,
described in Section 6.2) randomly spaced throughout each
10-m2 (approximately [~] 100-ft2]) area to be measured
and representative of the coated surface. The ve spot
measurements shall be made for each 10-m2 (~100-ft2) area
as follows:
6.3.1 For areas of coating not exceeding 30 m2 (~300 ft2)
in area, each 10-m2 (~100 ft2) (or increment thereof) area shall
be measured.
6.3.2 For areas of coating greater than 30 m2 (~300 ft2)
in area and not exceeding 100 m2 (~1,000 ft2) in area, three
10-m2 (~100-ft2) (or increment thereof) areas shall be arbitrarily
selected by the inspector and measured.
6.3.3 For areas of coating greater than 100 m2 (~1000 ft2)
in area, the rst 100 m2 (~1000 ft2) shall be measured as stated
in Section 6.3.2. For each additional 100-m2 (~1000-ft2) coated
area (or increment thereof), arbitrarily select and measure one
10-m2 (~100-ft2) area.
6.4 NONCONFORMING AREAS: If the coating thickness
for any 10-m2 (~100-ft2) area is not in compliance with the
contract documents, the procedure described here shall
be followed to assess the magnitude of the nonconforming
thickness.
6.4.1 Determine the spot DFT at 1.5-m (5-ft) intervals in
eight equally spaced directions radiating outward from the
nonconforming 10-m2 (~100-ft2) area. If there is no place to
measure in a given direction, then no measurement in that
direction is necessary. Acquire spot measurements in each
direction (up to the maximum surface area coated during the work
shift) until two consecutive conforming spot measurements are
acquired in that direction or until no additional measurements