9.3 Oxidize the specimen in the heat zone of a high-
temperature tube or muffle furnace open to the atmosphere at
825 625 °C. All specimens must be fully oxidized.
9.4 When the specimen has been completely oxidized (4 to
6 h), remove from the furnace and allow to cool.
NOTE 1—Complete oxidation of a sintered magnetic tungsten hard
metal sample can be checked by testing the cool oxidized chips with a
magnet. If any of the sample is still magnetic, recrush the sample and
place back in the furnace for further oxidation.
9.5 Pour the specimen onto a clean sheet of paper or into a
clean mortar and gently crush with a pestle.
9.6 Transfer the specimen to a labeled specimen vial.
9.7 Prepare a fusion mixture to be used with the oxidized
samples. For example, a mixture of dilithium tetraborate:
lithium metaborate with a 0.2 g addition of lithium bromide
can be used. The mixture should be used the day it is made.
Unused portions can be stored in an air tight container or
dessicator.
NOTE 2—Other fusion materials can be used. See 5.2.
9.8 In a fume hood, transfer the fusion mixture to a platinum
crucible immediately prior to weighing of the oxidized sample
material.
9.9 Weigh out 1.0000 60.0005 g of oxidized specimen and
transfer to the platinum crucible. Mix gently with the fusion
mixture.
NOTE 3—If there is not enough sample to make a standard fusion, or the
amount of the total mixture is too large for the casting dish, proportionate
amounts of oxidized test sample and fusion mixture can be utilized to
prepare a specimen recognizing that larger fractional errors may be
incurred in the analysis. This should only be used when absolutely
necessary.
9.10 Using the fluxer, melt the specimen at the lowest
temperature required for dissolution of the sample by the
fusion mixture used and cast into a heated platinum casting
dish.
9.10.1 Warning—The process of making glass fusions
exposes personnel to high-temperature liquids. Extreme care
should be exercised while preparing these samples. These high
temperatures also cause some volatilization of the lithium
compounds. The fluxer should have an exhaust hood to remove
these gases from the facility. The lithium compounds used in
this procedure are hygroscopic. Material open to the atmo-
sphere for an extended period of time will absorb moisture.
Exposure of this material to subsequent high heat will cause
rapid formation of steam and may cause spattering of the
molten glass onto the instrument and possibly the operator.
9.11 While the fused specimen is cooling, remove the
crucible from the instrument with the platinum-tipped tongs
and cool.
9.12 Place the crucible in a 1000 mL beaker which has a
2-volume percent solution of citric acid. Put the beaker on a hot
plate and warm the solution. The crucible should be clean in
approximately 30 min. Remove the crucible from the acid bath
with tongs and rinse with water. Dry the crucible and store.
9.13 When the fused specimen is cool, remove from the
casting dish by gripping the dish firmly with tongs, turning the
dish over, and gently tapping against a clean paper. The dish
and fused specimen should cleanly separate. Label the fused
specimen with a self-adhering tag.
NOTE 4—Any evidence of wetting between the specimen and the
platinum crucible or casting dish is an indication that the specimen has
reacted with these vessels and is not a valid representative sample.
9.14 If the fusion crystallizes (cooling slowly) or fractures
(cooling fast), crush the fusion and recast. If the fused
specimen cannot be removed from the platinum casting dish
with very light tapping, dissolve the specimen from the dish
using a warm 2-volume percent citric acid solution. Prepare a
new specimen in accordance with 9.7 – 9.10.
Caution—Excessive prying or tapping of the crystallized
specimen while it is in the dish will damage the platinum ware.
10. Standardization of Spectrometer and Analysis
10.1 Based on the X-ray spectrometer configuration and
instrument manufacturer’s operating instructions, determine
the instrument operating parameters to provide optimum spec-
tral analysis for each element being analyzed in a given matrix.
Table 2 provides the approximate X-ray peak positions (Bragg
angle - 2Θ) and crystals typically used for each of the elements
of interest.
10.2 If required, normalize the X-ray spectrometer operat-
ing parameters to obtain the appropriate secondary X-ray
intensities from the reference standards utilized.
10.3 Measure X-ray intensities on a sufficient number of
fused standards to establish a calibration curve (intensity
versus concentration of analyte) for each element of interest.
NOTE 5—The number of standards sufficient to establish a calibration
curve is dependent on the range of concentrations to be analyzed for each
element. In all cases, a minimum of six standards is required.
10.4 Calibration curves may be established manually, or
corrections for interelement effects may be calculated using
XRF vendor-supplied computer software.
NOTE 6—Accuracy of a given interelement correction routine can be
verified by including one or more reference standards as “blind” un-
knowns as part of an analysis.
11. Procedure
11.1 Obtain X-ray intensity data from the fused test speci-
mens.
TABLE 2 Analytical X-ray Lines
Element
Symbol
Shell
Series
Reflection
Order
Bragg
Angle 2Θ
Wavelength,
A
Crystal
Co Kα1 52.788 1.7906 LiF100
Cr Kα1 69.368 2.2913 LiF100
Fe Kα1 57.526 1.9376 LiF100
Hf Lα1 45.880 1.5690 LiF100
Mo Kα1 20.276 0.7092 LiF100
Nb Kα1 21.340 0.7461 LiF100
Ni Kα1 48.632 1.6594 LiF100
Ta Lα1 64.640 1.5222 LiF110
Ti Kα1 86.186 2.7502 LiF100
VKα1 123.172 2.5054 LiF110
B890 − 20
3