
Copyright © 2024 American Association of Textile Chemists and Colorists
AATCC Manual of International Test Methods and Procedures/2025 AATCC TM194-2023 387
coating should be avoided; as it can melt
and interfere with mite recovery during
the subsequent heat step. Tangle Trap has
also been found to provide an effective
barrier (see 12.3). However, final recov-
ery numbers may be negatively influ-
enced due to the numbers of mites be-
coming irreversibly trapped on the
material. A third option is to place a tex-
tile sample that has previously been dem-
onstrated to be effective as a mite barrier
fabric tightly across the top of the test
chamber and fix in place.
8.1.3 Close each test chamber and
place the test units at 25 ± 1°C (77 ± 1°F)
and 73-76% relative humidity for approx-
imately 48 h to acclimatize the specimen
microenvironment.
8.1.4 Place 25 male and 25 female
mites from a robust colony on each accli-
matized test and control specimen.
8.1.5 If possible, use mating pairs to
ensure that the females are at a similar
stage of oviposition.
8.1.6 Close the test chamber and incu-
bate the samples at 25 ± 1°C (77 ± 1°F)
and 73-76% relative humidity for six
weeks.
8.2 Mite Recovery.
8.2.1 After the six weeks of incubation,
remove one test chamber at a time from
the incubator. For each plate, pre-cut a
sample of nylon mesh material (see 12.4)
to fit snugly within the test plate. Cover
the mesh with adhesive film so that the
sticky side of the tape is in direct contact
with the mesh, and the edges of the tape
extends beyond the edge of the mesh by
approximately 5 mm.
8.2.2 Remove the lid from the test
chamber and firmly place the tape/mesh
combination directly on top of the test
specimen, with the sticky side of the tape
facing down. Ensure the outer edges of
the tape adhere to the sides of the test
plate. The mesh should serve to limit the
amount of food particles and dead mites
that might adhere to the sticky tape dur-
ing the subsequent recovery step.
8.2.3 Place the plate, minus the lid, di-
rectly on a heating plate (see 12.5) set at
50°C (122°F). Place a weight on top of
the tape/mesh combination to ensure inti-
mate contact between the mesh and the
test specimen. Use of a pre-cut circle of
Styrofoam that will fit in the test plate be-
tween the tape/mesh and the weight can
aid in evenly distributing the weight, and
may also limit moisture buildup on the
tape that might otherwise occur.
8.2.4 Leave each test chamber on the
hot plate for a minimum of 5 h. This
should be sufficient for mite recovery
from textiles, as well as thicker samples
or those with a heavy backing, such as
carpets.
8.2.5 After the heat exposure, remove
the weight and recover the tape/mesh.
Cover the second side of the mesh with a
clear polyethylene film or a second coat
of sticky tape, to secure the mites for sub-
sequent counting.
9. Evaluation
9.1 For each specimen, count the total
number of mites recovered on the film
using a low power stereo-binocular micro-
scope.
9.2 For each set of test and control
specimens, calculate the mean and stan-
dard deviation.
10. Report
10.1 Express results as percent reduc-
tion versus the control, using the follow-
ing formula:
where:
R= percent reduction of the test spec-
imen versus the control.
A= the mean number of dust mites
found on the control specimen.
B= the mean number of dust mites
found on the test specimen.
10.2 Indicate that test was conducted
following AATCC TM194-2023.
10.3 For a valid test: If a healthy mite
colony has been established on the nega-
tive control specimens during the six
week test period, mites on the internal lab
control do not need to be recovered and
counted. However, if the mite numbers
on the negative controls are lower than
expected for a control in this time period,
mites from the internal lab control must
be recovered as outlined in 8.2. Recovery
numbers outside of the normal range of
mites previously established for each of
the internal controls, invalidate the test.
10.4 Any deviation from this proce-
dure as written must be reported.
10.5 The criterion for passing the test
must be determined by the interested parties.
11. Precision and Bias
11.1 Precision. Precision for this test
method has not been established. Until a
precision statement is generated for this
test method, use standard statistical tech-
niques in making any comparisons of test
results for either within-laboratory or be-
tween-laboratory averages.
11.2 Bias. Anti-house dust mite prop-
erties of textiles under long-term test con-
ditions can be defined only in terms of a
test method. There is no independent
method for determining the true values.
As a means of estimating these proper-
ties, the method has no known bias.
12. Notes
12.1 Available from Publication Office,
ACGIH, Kemper Woods Center, 1330 Kemper
Meadow Dr., Cincinnati OH 45240; tel: +1.
513.742.2020; web site: www.acgih.org.
12.2 TetraMin Tropical Flakes con be ob-
tained from Spectrum Brands Pet LLC 3001
Commerce St., Blacksburg VA 24060-6617.
phone 800.526.0650 and via their website at
https://www.tetra-fish.com or other online re-
tailers.
12.3 Tangle Trap can be obtained from The
Tanglefoot Company, 314 Straight Ave. S.W.,
Grand Rapids MI 49504-6485; tel: +1.616.
459.4139; fax: +1.616.459.4139; e-mail: info@
tanglefoot.com; web site: www.tanglefoot.com.
12.4 A mesh consisting of 447 denier nylon
filaments, with 40 × 40 threads/inch, thickness
of 0.36 mm and with a hole size of 0.63 mm
has been found effective. This mesh can be
obtained from Industrial Textiles Ltd., 62 Pa-
tiki Rd., Avondale, Auckland 1007 NZ; tel:
+64.98283188; free fax: +64.98281022;
email: info@vakeattack.co.nz; web site: www.
indtex.co.nz.
12.5 Other means of heating the samples,
such as the use of an incandescent light
source, can also be used. However, the process
must be optimized to obtain maximum mite
recovery for each type of sample before initia-
tion of the test.
13. History
13.1 Revised in 2023 to add dust mite food
source and to align to the AATCC Style guide-
lines.
13.2 Editorially revised 2019; reaffirmed
2013; editorially revised 2010; reaffirmed
2008, 2007.
13.3 Developed 2006 by AATCC Committee
RA49.
Copyright AATCC International
Provided by Accuris under license with AATCC
No reproduction or networking permitted without license from Accuris Not for Resale, 2025-04-14 18:24:10 UTC
Sold to: Nawal Analytical Labs India Private limited
Order Number: 00015024