8.3 Oxidant, air, oil free. (Warning—Compressed gas un-
der high pressure.)
8.4 Make-Up Gas, nitrogen, 99.99 % pure. (Warning—
Compressed gas under high pressure.)
8.5 Reference Standards:
8.5.1 tert-Amyl methyl ether,
4,5
(Warning—Flammable
liquid. Harmful if inhaled.)
8.5.2 Butane, (Warning—Flammable liquid. Harmful if
inhaled.)
8.5.3 tert-Butanol, (Warning—Flammable liquid. Harmful
if inhaled.)
8.5.4 sec-Butyl methyl ether,
6,5
(Warning—Flammable
liquid. Harmful if inhaled.)
8.5.5 4,4-Dimethyl-2-neopentyl-1-pentene,
7,5
(Warning—
Flammable liquid. Harmful if inhaled.)
8.5.6 Isobutylene, (Warning—Flammable liquid. Harmful
if inhaled.)
8.5.7 Methanol, (Warning—See Note 1.)
NOTE 1—Warning: Toxic Flammable Liquid. Harmful if inhaled or
ingested.
8.5.8 2-Methyl-2-butene,
7,5
(Warning—Flammable liquid.
Harmful if inhaled.)
8.5.9 Methyl tert-butyl ether, 99 + % pure,
8,5
(Warning—
Flammable liquid. Harmful if inhaled.)
8.5.10 2,2,4,6,6-Pentamethyl-3-heptene,
7,5
(Warning—
Flammable liquid. Harmful if inhaled.)
8.5.11 n-Pentane, (Warning—Flammable liquid. Harmful
if inhaled.)
8.5.12 cis-2-Pentene, (Warning—Flammable liquid. Harm-
ful if inhaled.)
8.5.13 trans-2-Pentene, (Warning—Flammable liquid.
Harmful if inhaled.)
8.5.14 2,4,4-Trimethyl-1-pentene, (Warning—Flammable
liquid. Harmful if inhaled.)
8.5.15 2,4,4-Trimethyl-2-pentene, (Warning—Flammable
liquid. Harmful if inhaled.)
8.5.16 1 % Contaminant Standard,
9,5
contains 1.0 % of
some of the contaminants in MTBE, (Warning—Flammable
liquid. Harmful if inhaled.)
8.5.17 0.1 % Contaminant Standard,
9,5
contains 0.1 % of
some of the contaminants in MTBE, (Warning—Flammable
liquid. Harmful if inhaled.)
9. Sampling
9.1 MTBE can be sampled either in a floating piston
cylinder or into an open container since vapor pressures less
than 70 kPa (10 psi) are expected.
9.1.1 Cylinder Sampling—Refer to Practice D3700 for in-
structions on transferring a representative sample from a source
into a floating piston cylinder. Add inert gas to the ballast side
of the piston to achieve a pressure of 310 kPa (45 psi) above
the vapor pressure of the sample.
9.1.2 Open Container Sampling—Refer to Practice D4057
for instructions on manual sampling from bulk storage into
open containers. Stopper container immediately after drawing
sample.
9.2 Preserve the sample by cooling to approximately 4 °C
and by maintaining that temperature until immediately prior to
analysis.
9.3 Transfer an aliquot of the cooled sample into a pre-
cooled septum vial, then seal appropriately. Obtain the test
specimen for analysis directly from the sealed septum vial, for
either manual or automatic syringe injection.
10. Preparation of Apparatus
10.1 Install and condition column in accordance with manu-
facturer’s or supplier’s instructions. After conditioning, attach
column outlet to flame ionization detector inlet and check for
leaks throughout the system. When leaks are found, tighten or
replace fittings before proceeding.
10.2 Adjust the carrier gas flow rate so that an average
linear velocity at the starting temperature of the run is between
21 cm ⁄s and 24 cm ⁄s, as determined in Eq 1. Flow rate
adjustment is made by raising or lowering the carrier gas
pressure (head pressure) to the injector. The following starting
point pressures can be useful to adjust the carrier gas flow:
Column length 50 m 100 m 150 m
Starting point pressure, kPa (psig) 262 (38) 275 (40) 552 (80)
10.2.1 Average Linear Gas Velocity:
u
ave
5L/t
m
(1)
where:
L= the length of the column in cm, and
t
m
= the retention time in seconds of methane.
10.3 Adjust the operating conditions of the gas chromato-
graph to conform to the list in Table 1. Turn on the detector,
ignite the flame, and allow the system to equilibrate.
10.4 When the method is first set up, ensure that the FID is
not saturated. Plot the peak area versus MTBE concentration
for prepared standards in the concentration range of interest. If
the plot is not linear, increase the split ratio, or use a less
sensitive detector range, or both.
11. Column Evaluation and Optimization
11.1 In order to establish that the column/temperature pro-
gram will perform the required separation, the resolution
between cis-2-pentene and tert-butanol and between trans-2-
pentene and tert-butanol must be determined. The retention of
4
The sole source of supply of the apparatus known to the committee at this time
is Sigma-Aldrich Chemical Company, Inc., Milwaukee, WI. A 96 % pure sample
obtained from Sigma-Aldrich was the highest purity found.
5
If you are aware of alternative suppliers, please provide this information to
ASTM International Headquarters. Your comments will receive careful consider-
ation at a meeting of the responsible technical committee,
1
which you may attend.
6
The sole source of supply of the apparatus known to the committee at this time
is Farcham Laboratories, Gainesville, FL.
7
The sole source of supply of the apparatus known to the committee at this time
is Organic Technologies (formerly Wiley Organics), P.O. Box 640, 1245 S. 6th St.,
Coshocton, OH 43812.
8
The sole source of supply of the apparatus, HPLC grade MTBE, known to the
committee at this time is from Aldrich Chemical Company, Inc., Milwaukee, WI.
9
The sole source of supply of the apparatus, reference samples that contain only
contaminants boiling above ambient, known to the committee at this time is
Supelco, Inc., Bellefonte, PA.
D5441 − 21 (2025)
3