Designation: E353 −25
Standard Test Methods for
Chemical Analysis of Stainless, Heat-Resisting, Maraging,
and Other Similar Chromium-Nickel-Iron Alloys
1
This standard is issued under the fixed designation E353; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 These test methods cover the chemical analysis of
stainless, heat-resisting, maraging, and other similar
chromium-nickel-iron alloys having chemical compositions
within the following limits:
Element Composition Range, %
Aluminum 0.002 to 5.50
Boron 0.001 to 0.20
Carbon 0.01 to 1.50
Chromium 0.01 to 35.00
Cobalt 0.01 to 15.00
Niobium 0.01 to 4.00
Copper 0.01 to 5.00
Lead 0.001 to 0.50
Manganese 0.01 to 20.00
Molybdenum 0.01 to 7.00
Nickel 0.01 to 48.00
Nitrogen 0.001 to 0.50
Phosphorus 0.002 to 0.35
Selenium 0.01 to 0.50
Silicon 0.01 to 4.00
Sulfur 0.002 to 0.50
Tantalum 0.01 to 0.80
Tin 0.001 to 0.05
Titanium 0.01 to 4.50
Tungsten 0.01 to 4.50
Vanadium 0.005 to 1.00
Zirconium 0.001 to 0.20
1.2 The test methods in this standard are contained in the
sections indicated below:
Sections
Aluminum, Total, by the 8-Quinolinol Gravimetric
Method (0.20 % to 7.00 %)
119–126
Aluminum, Total, by the 8-Quinolinol Spectrophotometric
Method (0.003 % to 0.20 %)
71–81
Carbon, Total, by the Combustion–Thermal Conductivity
Method–Discontinued 1986
153–163
Carbon, Total, by the Combustion Gravimetric Method
(0.05 % to 1.50 %)–Discontinued 2013
98–108
Chromium by the Atomic Absorption Spectrometry
Method (0.006 % to 1.00 %)
202–211
Sections
Chromium by the Peroxydisulfate Oxidation–Titration
Method (0.10 % to 35.00 %)
212–220
Chromium by the Peroxydisulfate-Oxidation Titrimetric
Method-Discontinued 1980
145–152
Cobalt by the Ion-Exchange–Potentiometric Titration
Method (2 % to 15 %)
53–60
Cobalt by the Nitroso-R-Salt Spectrophotometric
Method (0.01 % to 5.0 %)
61–70
Copper by the Neocuproine Spectrophotometric Method
(0.01 % to 5.00) %)
109–118
Copper by the Sulfide Precipitation-Electrodeposition
Gravimetric Method (0.01 % to 5.00 %)
82–89
Lead by the Ion-Exchange-Atomic Absorption
Spectrometry Method (0.001 % to 0.50 %)
127–136
Manganese by the Periodate Spectrophotometric
Method (0.01 % to 5.00 %)
9–18
Molybdenum by the Ion Exchange–8-Hydroxyquinoline
Gravimetric Method
242–249
Molybdenum by the Thiocyanate Spectrophotometric
Method (0.01 % to 1.50 %)
190–201
Nickel by the Dimethylglyoxime Gravimetric Method (0.1
% to 48.0 %)
172–179
Phosphorus by the Alkalimetric Method (0.02 % to
0.35 %)
164–171
Phosphorus by the Molybdenum Blue
Spectrophotometric Method (0.002 % to 0.35 %)
19–30
Silicon by the Gravimetric Method (0.05 % to 4.00 %) 46–52
Sulfur by the Gravimetric Method-Discontinued 1988 30–36
Sulfur by the Combustion-Iodate Titration Method (0.005
% to 0.5 %)-Discontinued 2014
37–45
Sulfur by the Chromatographic Gravimetric Method-
Discontinued 1980
137–144
Tin by the Solvent Extraction–Atomic Absorption
Spectrometry Method (0.002 % to 0.10 %)
180–189
Tin by the Sulfide Precipitation-Iodometric Titration
Method (0.01 % to 0.05 %)
90–97
Titanium by the Diantipyrylmethane Spectrophotometric
Method (0.01 % to 0.35 %)
231–241
Vanadium by the Atomic Absorption Spectrometry
Method (0.006 % to 0.15 %)
221–230
1.3 Test methods for the determination of carbon and sulfur
not included in this standard can be found in Test Methods
E1019.
1.4 Some of the composition ranges given in 1.1 are too
broad to be covered by a single test method and therefore this
standard contains multiple test methods for some elements.
The user must select the proper test method by matching the
information given in the Scope and Interference sections of
each method with the composition of the alloy to be analyzed.
1
These test methods are under the jurisdiction of ASTM Committee E01 on
Analytical Chemistry for Metals, Ores, and Related Materials and are the direct
responsibility of Subcommittee E01.01 on Iron, Steel, and Ferroalloys.
Current edition approved May 15, 2025. Published May 2025. Originally
approved in 1968. Last previous edition approved in 2019 as E353 – 19
ɛ1
. DOI:
10.1520/E0353-25.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
1
北极星标准文库-高清标准规范分享平台 www.bjxwk.com
1.5 The values stated in SI units are to be regarded as
standard.
1.6 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
priate safety, health, and environmental practices and deter-
mine the applicability of regulatory limitations prior to use.
Specific hazards statements are given in Section 6and in
special “Warning” paragraphs throughout these test methods.
1.7 This international standard was developed in accor-
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the
Development of International Standards, Guides and Recom-
mendations issued by the World Trade Organization Technical
Barriers to Trade (TBT) Committee.
2. Referenced Documents
2.1 ASTM Standards:
2
E29 Practice for Using Significant Digits in Test Data to
Determine Conformance with Specifications
E50 Practices for Apparatus, Reagents, and Safety Consid-
erations for Chemical Analysis of Metals, Ores, and
Related Materials
E60 Practice for Analysis of Metals, Ores, and Related
Materials by Spectrophotometry
E135 Terminology Relating to Analytical Chemistry for
Metals, Ores, and Related Materials
E173 Practice for Conducting Interlaboratory Studies of
Methods for Chemical Analysis of Metals (Withdrawn
1997)
3
E350 Test Methods for Chemical Analysis of Carbon Steel,
Low-Alloy Steel, Silicon Electrical Steel, Ingot Iron, and
Wrought Iron
E351 Test Methods for Chemical Analysis of Cast Iron—All
Types
E352 Test Methods for Chemical Analysis of Tool Steels and
Other Similar Medium- and High-Alloy Steels
E354 Test Methods for Chemical Analysis of High-
Temperature, Electrical, Magnetic, and Other Similar Iron,
Nickel, and Cobalt Alloys
E882 Guide for Accountability and Quality Control in the
Chemical Analysis Laboratory
E1019 Test Methods for Determination of Carbon, Sulfur,
Nitrogen, and Oxygen in Steel, Iron, Nickel, and Cobalt
Alloys by Combustion and Inert Gas Fusion Techniques
E1024 Guide for Chemical Analysis of Metals and Metal
Bearing Ores by Flame Atomic Absorption Spectropho-
tometry (Withdrawn 2004)
3
E1601 Practice for Conducting an Interlaboratory Study to
Evaluate the Performance of an Analytical Method
E1806 Practice for Sampling Steel and Iron for Determina-
tion of Chemical Composition
2.2 Other Document:
ISO 5725 Precision of Test Methods—Determination of
Repeatability and Reproducibility for Inter-Laboratory
Tests
4
3. Terminology
3.1 For definitions of terms used in these test methods, refer
to Terminology E135.
4. Significance and Use
4.1 These test methods for the chemical analysis of metals
and alloys are primarily intended as referee methods to test
such materials for compliance with compositional
specifications, particularly those under the jurisdiction of
ASTM Committee A01 on Steel, Stainless Steel, and Related
Alloys. It is assumed that all who use these test methods will
be trained analysts capable of performing common laboratory
procedures skillfully and safely. It is expected that work will be
performed in a properly equipped laboratory under appropriate
quality control practices such as those described in Guide
E882.
5. Apparatus, Reagents, and Instrumental Practices
5.1 Apparatus—Specialized apparatus requirements are
listed in the “Apparatus” Section in each method.
5.1.1 In the methods specifying spectrophotometric testing,
the cells utilized to contain the reference material solutions and
the sample solutions in spectrophotometers are referred to as
“absorption cells.” Please note that the radiant energy passed
through the cells can be measured as absorbance or transmit-
tance. These methods refer to absorbance measurements. Refer
to Practices E60 for details.
5.2 Reagents:
5.2.1 Purity of Reagents—Refer to Practice E50, Section
5.2, for the purity of reagents provision.
5.2.2 Purity of Water—Refer to Practice E50, Section 5.1,
for the purity of water provision.
5.3 Spectrophotometric Practice—Spectrophotometric
practice prescribed in these test methods shall conform to
Practice E60.
6. Hazards
6.1 For precautions to be observed in the use of certain
reagents and equipment in these methods, refer to Practices
E50.
7. Sampling
7.1 For procedures to sample the material, refer to Practice
E1806.
8. Interlaboratory Studies and Rounding Calculated
Values
8.1 These test methods have been evaluated as directed in
Practice E173 (withdrawn 1997) or ISO 5725. Practice E173
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at www.astm.org/contact. For Annual Book of
ASTM Standards volume information, refer to the standard’s Document Summary
page on the ASTM website.
3
The last approved version of this historical standard is referenced on
www.astm.org.
4
Available from American National Standards Institute (ANSI), 25 W. 43rd St.,
4th Floor, New York, NY 10036, www.ansi.org.
E353 − 25
2
北极星标准文库-高清标准规范分享平台 www.bjxwk.com
has been replaced by Practice E1601. The reproducibility, R2,
of Practice E173 corresponds to the reproducibility index Rof
Practice E1601. The repeatability, R1, of Practice E173 corre-
sponds to the repeatability index, r, of Practice E1601.
8.2 Rounding of test results obtained using this test method
shall be performed as directed in ASTM E29, Rounding
Method, unless an alternative rounding method is specified by
the customer or applicable material specification.
MANGANESE BY THE META PERIODATE
SPECTROPHOTOMETRIC METHOD
9. Scope
9.1 This method covers the determination of manganese
from 0.01 % to 5.00 %.
10. Summary of Method
10.1 Manganous ions are oxidized to permanganate ions by
treatment with periodate. Tungsten when present at composi-
tions greater than 0.5 % is kept in solution with H
3
PO
4
.
Solutions of the samples are fumed with HClO
4
so that the
effect of periodate is limited to the oxidation of manganese.
Spectrophotometric absorbance measurement is made at 545
nm.
11. Concentration Range
11.1 The recommended concentration range is 0.15 mg to
0.8 mg of manganese per 50 mL of solution, using a 1-cm cell
(see Note 1) and a spectrophotometer with a band width of 10
nm or less.
NOTE 1—This method has been written for cells having a 1-cm light
path and a “narrow-band” instrument. The concentration range depends
upon band width and spectral region used as well as cell optical path
length. Cells having other dimensions may be used, provided suitable
adjustments can be made in the amounts of sample and reagents used.
12. Stability of Color
12.1 The color is stable for at least 24 h.
13. Interferences
13.1 HClO
4
treatment, which is used in the procedure,
yields solutions which can be highly colored due to the
presence of Cr (VI) ions. Although these ions and other colored
ions in the sample solution undergo no further change in color
quality upon treatment with metaperiodate ion, the following
precautions must be observed when filter spectrophotometers
are used: Select a filter with maximum transmittance between
545 nm and 565 nm. The filter must transmit not more than 5 %
of its maximum at a wavelength shorter than 530 nm. The band
width of the filter should be less than 30 nm when measured at
50 % of its maximum transmittance. Similar restrictions apply
with respect to the wavelength region employed when other
“wide-band” instruments are used.
13.2 The spectral transmittance curve of permanganate ions
exhibits two useful minima, one at approximately 526 nm, and
the other at 545 nm. The latter is recommended when a
“narrow-band” spectrophotometer is used.
13.3 Tungsten, when present in amounts of more than 0.5 %
interferes by producing a turbidity in the final solution. A
special procedure is provided for use with samples containing
more than 0.5 % tungsten which eliminates the problem by
preventing the precipitation of the tungsten.
14. Reagents
14.1 Manganese, Standard Solution (1 mL = 0.032 mg
Mn)—Transfer the equivalent of 0.4000 g of assayed, high-
purity manganese (purity: 99.99 % minimum), to a 500-mL
volumetric flask and dissolve in 20 mL of HNO
3
by heating.
Cool, dilute to volume, and mix. Using a pipet, transfer 20 mL
to a 500-mL volumetric flask, dilute to volume, and mix.
14.2 Nitric-Phosphoric Acid Mixture—Cautiously, while
stirring, add 100 mL of HNO
3
and 400 mL of H
3
PO
4
to 400
mL of water. Cool, dilute to 1 L, and mix. Prepare fresh as
needed.
14.3 Potassium Metaperiodate Solution (7.5 g/L)—Dissolve
7.5 g of potassium metaperiodate (KIO
4
) in 200 mL of hot
HNO
3
(1 + 1), add 400 mL of H
3
PO
4
, cool, dilute to 1 L, and
mix.
14.4 Water, Pretreated with Metaperiodate—Add 20 mL of
KIO
4
solution to 1 L of water, mix, heat at not less than 90°C
for 20 min to 30 min, and cool. Use this water to dilute
solutions to volume that have been treated with KIO
4
solution
to oxidize manganese, and thus avoid reduction of permangan-
ate ions by any reducing agents in the untreated water.
Caution—Avoid the use of this water for other purposes.
15. Preparation of Calibration Curve
15.1 Calibration Solutions—Using pipets, transfer 5 mL, 10
mL, 15 mL, 20 mL, and 25 mL of manganese standard solution
(1 mL = 0.032 mg Mn) to 50-mL borosilicate glass volumetric
flasks, and if necessary, dilute to approximately 25 mL.
Proceed as directed in 15.3.
15.2 Reference Solution—Transfer approximately 25 mL of
water to a 50-mL borosilicate glass volumetric flask. Proceed
as directed in 15.3.
15.3 Color Development—Add 10 mL of KIO
4
solution,
and heat the solutions at not less than 90°C for 20 min to 30
min (Note 2). Cool, dilute to volume with pretreated water, and
mix.
NOTE 2—Immersing the flasks in a boiling water bath is a preferred
means of heating them for the specified period to ensure complete color
development.
15.4 Spectrophotometry:
15.4.1 Multiple-Cell Spectrophotometer—Measure the cell
correction using the Reference Solution (15.2) in absorption
cells with a 1-cm light path and using a light band centered at
545 nm. Using the test cell, take the spectrophotometric
absorbance readings of the calibration solutions versus the
Reference Solution (15.2).
15.4.2 Single-Cell Spectrophotometer—Transfer a suitable
portion of the Reference Solution (15.2) to an absorption cell
with a 1-cm light path and adjust the spectrophotometer to the
initial setting, using a light band centered at 545 nm. While
E353 − 25
3
北极星标准文库-高清标准规范分享平台 www.bjxwk.com

标签: #不锈钢 #钢 #合金

摘要:

ASTM E353 - 25 是国际权威材料测试协会发布的最新版本标准,专门针对不锈钢、耐热钢、马氏体时效钢及其他类似铬镍铁合金的化学成分分析提供标准试验方法。该标准详细规定了如何通过湿化学法、仪器分析法等手段准确测定钢中主要合金元素与微量元素的含量,确保材料符合工业制造、航空航天、石化设备及医疗器械等高端应用领域的质量要求。遵循ASTM E353 - 25可帮助企业有效控制原料与成品成分偏差,提升检测数据的国际互认性,优化产品质量并满足欧美市场准入标准。本文件适用于冶金实验室、第三方检测机构及

展开>> 收起<<
ASTM E353 - 25 不锈钢、耐热钢、马氏体时效钢及其他类似铬镍铁合金化学分析的标准试验方法.pdf

共46页,预览3页

还剩页未读, 继续阅读

声明:本文档系会员上传,若文档所含内容侵犯了您的版权或隐私,请立即通知,我们立即给予侵权申诉删除!
作者:标准小能手 分类:国外协会 价格:20星币 属性:46 页 大小:2.02MB 格式:PDF 时间:2026-06-17

开通VIP享超值会员特权

  • 多端同步记录
  • 高速下载文档
  • 免费文档工具
  • 分享文档赚钱
  • 每日登录抽奖
  • 优质衍生服务
/ 46
客服
关注