standards. For DCPD, the resulting second derivative peak near
1610 cm
−1
is related to DCPD mass fraction by calibrating the
instrument with known EPDM standards.
3.3 Two main steps are performed in this procedure: auto-
matic determination of film thickness and quantification of
diene.
3.4 For oil-extended polymers, the oil must be extracted
before diene is determined. Test Methods D1416, Sections 67
through 74 and Test Methods D297 can be used for this
purpose.
4. Apparatus
4.1 Press, Carver-type, capable of pressing films at 150°C
and 10 MPa or higher.
4.1.1 Primary Mold—A stainless steel (SS) mold, approxi-
mately 400 µm thick, with an opening sized appropriately for
the specimen film holder described in 4.4 (typically 2 by 2 cm).
The mold should have approximately the same dimensions as
the press platens.
4.1.2 Alternate Mold—A thinner mold may be used;
however, precision may be adversely affected at low diene
levels. Test precision should be determined when thinner films
are utilized. Example: a 127-µm (0.005-in) SS shim with a 15-
by 35-mm opening may be used for the simultaneous determi-
nation of ethylene (Test Methods D3900) and diene.
4.1.3 Alternative Mold for Low Molecular Weight (Liquid)
EPDM Terpolymers—For liquid EPDM a ring washer 22 mm
OD by 16 mm ID (7/8 in OD by 5/8 inch in ID) 400 µm thick
is used as a apacer between salt plates (NaBr, NaCl) to set a
fixed path length. The spacer is sized to cover only the outer
edge of the salt plate.
4.2 Sheets for Molding, either PTFE-coated aluminum
sheets or Mylar, Type A, 36 µm thick or less.
4.3 Specimen Film Holders—Films may be molded, cut, and
transferred to a film holder. Magnetic film holders are ideal.
Alternatively, a mold sized to fit in the spectrometer specimen
compartment, with an appropriate size opening may be used to
support the film without removal after pressing.
4.4 Infrared Fourier Transform Spectrophotometer—An in-
strument capable of measuring absorbances in the range 400 to
6000 cm
−1
with a transmittance specification (accuracy) of 6
1 % Tor better. The instrument should be capable of spectral
resolution of 2 cm
−1
(see “Specification for Evaluation of
Research Quality Analysis of Infrared Spectra”). A deuterated
triglycine sulfate (DTGS) detector should be used. The instru-
ment must be operating in accordance with Practices E168.
Practices E932 and E1421 are other important references.
4.4.1 The instrument should be capable of spectral
accumulation, averaging, and subtracting capabilities. Water is
the primary source of interference in this method. Methods,
physical and electronic, that minimize moisture level and
variation are required to obtain the highest precision. The
preferred method is use of an instrument equipped with a dry
gas purge and a specimen shuttle, which permits alternating
and repetitive collection of single beam specimen and back-
ground spectra (see Section 6). Alternatively, should a speci-
men shuttle be unavailable, careful purging of the sample
compartment with dry nitrogen can yield satisfactory results.
High precision of calibration standard data is indicative of
adequate purging. When moisture interference is not removed
by purging, spectral subtraction of water vapor may be used. A
procedure for further method development is described in
Appendix X1.
5. Test Specimen Preparation
5.1 Primary Method—Place an appropriate amount of the
test specimen to fill the mold opening (typically 0.2 to 0.5 g)
between two PTFE-coated aluminum or Mylar sheets in the
mold. Place the mold between the press platens heated to 125
65°C and apply approximately 10 MPa or higher pressure for
60 610 s.
NOTE 2—High molecular weight polymers may require higher tempera-
tures and pressures to obtain a good test specimen.
5.1.1 Cool the specimen to ambient temperature. Cut, if
necessary, a piece of film to the appropriate size to cover the
specimen holder window. Detach the specimen film from the
aluminum or Mylar and position it on the spectrophotometer
sample holder window.
5.2 Alternate Method—When using the thinner mold de-
scribed in 4.1.2, place a small piece (0.04–0.06 g) of test
specimen in the mold opening between two Mylar sheets and
press as in 5.1. Remove the mold from the press, turn it over
and press it again, then remove the mold from the press and
cool it to ambient temperature. When cool, carefully remove
the Mylar sheets, allowing the specimen film to remain
attached to the mold.
5.3 Alternate Method for Liquid Sample Film Preparation—
Place a washer (described in 4.1.3) on top of a salt plate. Place
a small amount (about 0.3 g) of liquid EPDM polymer in the
center of the washer filling the hole completely. Place a second
salt plate on top of the filled washer. Gently place a 1 kg weight
on top of the salt plate/washer assembly and allow the
weighted assembly to sit for 2 to 3 min. (For viscous samples
it may be necessary to warm the sample prior to pressing.)
Remove the weight and allow to cool, if necessary. Wipe off
any excess material that may have been pressed out of the
assembly. Hold the assembly up to the light and inspect for
bubbles or voids, or both. Should there be imperfections, repeat
the sample preparation with a larger amount of sample.
6. Spectral Acquisition
6.1 With a specimen shuttle:
6.1.1 Data acquisition parameters:
6.1.1.1 Resolution: 2 cm
−1
.
6.1.1.2 Scans/Scan time: Total scan time required, split
between specimen and background, is about 90 s.
6.1.2 Place the test specimen in the specimen compartment,
allow purge to reestablish, and in alternating fashion, collect
single beam specimen (P) and “empty specimen compartment”
(P
0
) spectra. Eight passes of the shuttle should be sufficient
(eight specimen and eight empty compartment collections),
collecting four scans at each position.
D6047 − 17 (2025)
2