Designation: E1530 −25
Standard Test Method for
Evaluating the Resistance to Thermal Transmission by the
Guarded Heat Flow Meter Technique
1
This standard is issued under the fixed designation E1530; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
1. Scope*
1.1 This test method describes a steady-state technique for
the determination of the resistance to thermal transmission
(thermal resistance) of materials having a thickness of less than
25 mm. Thermal conductivity may be determined for homoge-
neous opaque solid specimens (see Note 1). This test method is
particularly useful for homogeneous, multilayer, and compos-
ite specimens having a thermal resistance in the range from
10 (cm)
2
·K·W
-1
to 400 (cm)
2
·K·W
-1
, which may be obtained
from materials with an approximate thermal conductivity range
0.1 W·m
-1
·K
-1
to 30 W·m
-1
·K
-1
over the approximate tempera-
ture range from 150 K to 600 K. It can be used outside these
ranges with reduced accuracy for thicker specimens and for
thermal conductivity values up to 60 W·m
-1
·K
-1
.
NOTE 1—A body is considered homogeneous when the property to be
measured is found to be independent of specimen dimensions.
1.2 This test method is similar in concept to Test Method
C518, but is modified to accommodate smaller test specimens,
having a higher thermal conductance. In addition, significant
attention has been paid to ensure that the thermal resistance of
contacting surfaces is minimized and reproducible.
1.3 The values stated in SI units are to be regarded as
standard. No other units of measurement are included in this
standard.
1.4 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
priate safety, health, and environmental practices and deter-
mine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accor-
dance with internationally recognized principles on standard-
ization established in the Decision on Principles for the
Development of International Standards, Guides and Recom-
mendations issued by the World Trade Organization Technical
Barriers to Trade (TBT) Committee.
2. Related Documents
2.1 ASTM Standards:
2
C518 Test Method for Steady-State Thermal Transmission
Properties by Means of the Heat Flow Meter Apparatus
D4417 Test Methods for Field Measurement of Surface
Profile of Blast Cleaned Steel
E3 Guide for Preparation of Metallographic Specimens
3. Terminology
3.1 Definitions of Terms Specific to This Standard:
3.1.1 heat flux transducer (HFT), n—a device that produces
an electrical output that is a function of the heat flux, in a
predefined and reproducible manner.
3.1.2 thermal conductance (C), n—the time rate of heat flux
through a unit area of a body induced by unit temperature
difference between the body surfaces.
3.1.2.1 average temperature of a surface, n—the area-
weighted mean temperature of that surface.
3.1.2.2 average (mean) temperature of a specimen (disc
shaped), n—the mean value of the upper and lower face
temperatures.
3.1.3 thermal conductivity (λ)—(of a solid material), n—the
time rate of heat flow, under steady conditions, through unit
area, per unit temperature gradient in the direction perpendicu-
lar to the area:
3.1.3.1 apparent thermal conductivity, n—when other
modes of heat transfer through a material are present in
addition to conduction, the results of the measurements per-
formed in accordance with this test method will represent the
apparent or effective thermal conductivity for the material
tested.
3.1.4 thermal resistance (R), n—the reciprocal of thermal
conductance.
3.2 Symbols:
λ= thermal conductivity, W·m
-1
·K
-1
1
This test method is under the jurisdiction of ASTM Committee E37 on Thermal
Measurements and is the direct responsibility of Subcommittee E37.05 on Thermo-
physical Properties.
Current edition approved Jan. 1, 2025. Published January 2025. Originally
approved in 1993. Last previous edition approved in 2019 as E1530 – 19. DOI:
10.1520/E1530-25.
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer service at www.astm.org/contact. For Annual Book of
ASTM Standards volume information, refer to the standard’s Document Summary
page on the ASTM website.
*A Summary of Changes section appears at the end of this standard
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
1
C= thermal conductance, W·m
-2
·K
-1
R= thermal resistance, m
2
·K·W
-1
∆x= specimen thickness, mm
A= specimen cross-sectional area, m
2
Q= heat flow, W
φ= heat flux transducer output, mV
N= heat flux transducer calibration constant, W·m
-2
·mV
-1
Nφ= heat flux, W·m
2
∆T= temperature difference, °C
T
g
= temperature of guard heater, °C
T
u
= temperature of upper heater, °C
T
l
= temperature of lower heater, °C
T
1
= temperature of one surface of the specimen, °C
T
2
= temperature of the other surface of the specimen, °C
T
m
= mean temperature of the specimen, °C
s
= unknown specimen
r
= known calibration or reference specimen
o
= contacts
4. Summary of Test Method
4.1 A specimen and a heat flux transducer (HFT) are
sandwiched between two flat plates controlled at different
temperatures, within an insulated guard tube, to produce an
uniform heat flow. A reproducible load is applied to the test
stack by pneumatic or other means, to ensure that there is a
reproducible contact resistance between the specimen and plate
surfaces. A guard surrounds the test stack and is maintained at
a uniform mean temperature of the two plates, in order to
minimize lateral heat flow to and from the stack. On attaining
steady state conditions, the difference in temperature between
the surfaces contacting the specimen is measured with tem-
perature sensors embedded in the plate surfaces, together with
the electrical output of the HFT. This output (voltage) is
proportional to the heat flow through the specimen, the HFT
and the interfaces between the specimen and the apparatus. The
proportionality is obtained through prior calibration of the
system with specimens of known thermal resistance measured
under the same conditions, such that contact resistance at the
surfaces can be made reproducible.
5. Significance and Use
5.1 This test method describes a procedure to measure and
compare the thermal resistance properties of specimens (less
than 25 mm in thickness) under controlled conditions.
6. Apparatus
6.1 A schematic rendering of a typical apparatus is shown in
Fig. 1. The position of the HFT relative to the specimen is not
important (in may be on the hot or cold side) as the method is
based on the maintenance of uniform axial heat flow with
minimal radial heat loss or gains. It is up to the designer/user
of the application to choose upward, downward, or horizontal
heat flow, although downwards heat flow in a vertical stack is
most commonly used.
6.2 Key Components of a Typical Device (The numbers 1 to
22 in parentheses refer to Fig. 1):
6.2.1 The adjustable compressive force for the stack is to be
provided by either a regulated pneumatic or hydraulic cylinder
(1), dead weights or a spring loaded mechanism. In either case,
means shall be provided to ensure that the loading can be
varied and set to selected reproducible values.
FIG. 1 Key Components of a Typical Device
E1530 − 25
2
6.2.2 The loading force shall be transmitted to the stack
through a gimball joint (2) that allows up to 5° swivel in the
plane perpendicular to the axis of the stack.
6.2.3 A suitable insulator plate (3) separates the gimball
joint from the upper plate (4).
6.2.4 The upper plate (assumed to be the hot plate for the
purposes of this description) is equipped with a heater (5) and
control temperature sensor (6) adjacent to the heater, to
maintain the desired temperature. (Other means of producing
and maintaining temperature may also be used providing the
requirements in 6.3 are met.) The construction of the upper
plate is such as to ensure uniform heat distribution across the
face in contact with the specimen (8). A temperature sensor (7)
that defines the temperature of the interface on the plate side is
attached to this face (or embedded in close proximity to it) such
that it does not interfere with the specimen/plate interface.
6.2.5 The specimen (8) is in direct contact with the upper
plate on one side and an intermediate plate (9) on the other
side.
6.2.6 The intermediate plate (9) is an optional item. Its
purpose is to provide a highly conductive environment to the
second temperature sensor (10), to obtain an average tempera-
ture of the surface. If the temperature sensor (10) is embedded
into the face of the HFT, or other means are provided to define
the temperature of the surface facing the specimen, the use of
the intermediate plate is not mandatory.
6.2.7 The heat flux transducer (HFT) is a device that will
generate an electrical signal in proportion to the heat flux
across it. The level of output required (sensitivity) significantly
depends on the rest of the instrumentation used for its
evaluation. The overall performance of the HFT and its readout
instrumentation shall be such as to meet the requirements in
Section 13.
6.2.8 The lower plate (12) is constructed similarly to the
upper plate (4), except it is positioned as the mirror image.
Both plates shall have a uniform thickness with the surfaces in
contact with the specimen being flat and parallel and having a
surface roughness of less than 10 µm maximum.
6.2.9 An insulator plate (16) separates the lower plate (12)
from the heat sink (17). In case of using circulating fluid in
place of a heater/temperature sensor arrangement in the upper
or lower plates, or both, the heat sink may or may not be
present.
6.2.10 The entire stack is surrounded by a guard tube,
insulated from heat losses to the outside, with a cross section
that is not too different from that of the stacks (18) equipped
with a heater or cooling coils (19), or both, and a similar
control temperature sensor (20) to maintain it at the mean
temperature between the upper and lower plates. A small,
generally unfilled, gap separates the guard from the stack. For
instruments limited to operate in the ambient region, no guard
is necessary although the use of a draft shield is recommended.
NOTE 2—For materials in the higher thermal conductivity range, in
particular, it is permissible to use a very thin layer of a suitable high
conductivity contact medium on the contacting surfaces of the specimen
to further reduce the thermal resistance of the interface and promote
uniform thermal contact across the interface area. Calibration shall be
performed under similar conditions of use or non-use of the contact
medium.
NOTE 3—Useful information on interfacial materials used for this test
method is available from literature.
3
NOTE 4—The cross-sectional area and the shape of the specimen can be
any, however, circular and rectangular cross sections are most commonly
used. Minimum size is dictated by the magnitude of the disturbance
caused by thermal sensors in relation to the overall flux distribution. The
most common sizes are 25 mm and 50 mm disks.
6.2.11 The measurement system is preferably equipped with
suitable means (21) to measure the in situ thickness of the
specimen and provide a means (22) to limit compression when
testing soft materials.
NOTE 5—This requirement is also mandatory for testing materials that
soften while heated.
6.3 Requirements:
6.3.1 Temperature control of upper and lower plate is to be
60.1 °C or better.
6.3.2 Reproducible load of 0.28 MPa has been found to be
satisfactory for solid specimens. The minimum load shall not
be less than 0.07 MPa.
6.3.3 Temperature sensors are usually fine gage or small-
diameter sheath thermocouples, however, ultraminiature resis-
tance thermometers and linear thermistors may also be used.
6.3.4 Operating range of a device using a mean temperature
guard shall be limited to from −100 °C to 300 °C, when using
thermocouples as temperature sensors, and from −180 °C to
300 °C when platinum resistance thermometers are used.
Thermistors are normally present on more restricted allowable
temperature range of use.
7. Sampling and Conditioning
7.1 Cut representative test specimens from larger pieces of
the sample material or body.
7.2 Condition the cut specimens in accordance with the
requirements of the appropriate material specifications, if any.
8. Test Specimen
8.1 The specimen to be tested should be representative for
the sample material. The recommended specimen configura-
tions of 25 mm or 50 mm 60.25 mm diameter disks or
squares, shall have smooth flat and parallel surfaces, finish of
10 µm (that is, an average roughness (Ra) of 10 µm (see Test
Methods D4417)) or less where possible, such that a uniform
thickness within 60.025 mm is attained in the range from
0.5 mm to 25.4 mm. For testing specimens having a thickness
less than 0.5 mm, an alternative technique, described in Annex
A1, is required.
NOTE 6—Guide E3 can be consulted for practices to achieve a suitably
smooth surface of the test material.
9. Calibration
9.1 Calibration of the system shall be undertaken using a set
of reference specimens having a range of thermal resistance
3
Stacey, C., Sumpkin, A. J., and Jarrett, R. N., “Techniques for Reducing
Thermal Contact Resistance in Steady-State Thermal Conductivity Measurements
on Polymer Composites,” International Journal of Thermophysics, Vol 37, 2016, p.
107.
E1530 − 25
3
摘要:

ASTM E1530-25是评估材料隔热性能的重要标准试验方法,采用防护热流计技术(Guarded Heat Flow Meter Technique)精确测量热传导率与热阻值。该标准适用于各种绝热材料、建筑保温层、工业隔热系统、冷热管道包覆层以及电子元器件散热材料的性能验证。通过严格控制的稳态热流条件下测量样品两侧温差与通过样品的热通量,ASTM E1530-25能够为工程师、质检人员和科研机构提供可比、可靠且可重复的隔热数据,帮助优化产品设计、满足建筑节能法规要求以及提升能效表现。该标准版本

展开>> 收起<<
ASTM E1530-25 用防护热流计技术评估隔热性能的标准试验方法.pdf

共8页,预览3页

还剩页未读, 继续阅读

声明:本文档系会员上传,若文档所含内容侵犯了您的版权或隐私,请立即通知,我们立即给予侵权申诉删除!
作者:老杨树 分类:国外协会 价格:20星币 属性:8 页 大小:283.26KB 格式:PDF 时间:2025-04-03

开通VIP享超值会员特权

  • 多端同步记录
  • 高速下载文档
  • 免费文档工具
  • 分享文档赚钱
  • 每日登录抽奖
  • 优质衍生服务
/ 8
客服
关注