Standard Test Method for
Determining the Particle Size Distribution of Advanced
Ceramics by Centrifugal Photosedimentation
1
This standard is issued under the fixed designation C 1282; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.
1. Scope
1.1 This test method covers determination of the particle
size distribution of advanced ceramic powders specifically
silicon nitride and carbides, in the range of 0.1 to 20 µm,
having a median particle diameter from 0.5 to 5.0 µm.
1.2 The procedure described in this test method may be
applied successfully to other ceramic powders in this general
size range, provided that appropriate dispersion procedures are
developed. It is the responsibility of the user to determine the
applicability of this test method to other materials.
1.3 The values stated in SI units are to be regarded as the
standard. The values given in parentheses are for information
only.
1.4 This standard does not purport to address all of the
safety concerns, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
2
C 242 Terminology of Ceramic Whitewares and Related
Products
3. Terminology
3.1 Definitions—Refer to Terminology C 242 for definitions
of terms used in this test method.
4. Summary of Test Method
4.1 A homogeneous aqueous dispersion of the powder is
prepared. While kept in a thoroughly mixed condition, a small
aliquot is transferred to the analyzer sample cell, which is
placed in the instrument and subjected to a controlled centrifu-
gal acceleration at a known or controlled temperature. At
predetermined times related to the sedimentation of specific
Stokes’ diameters (Note 1), the optical absorbance is recorded
and compared to the initial value to determine the fraction of
the total sample that has sedimented a specific distance. A
volume-based size distribution is calculated from the
absorbance-time data. The results are reported as equivalent
diameter (spherical) (Note 2) since particles in these powders
are not truly spherical.
NOTE 1—This diameter in µm is referred to as Din the equation given
below.
NOTE 2—Refer to Terminology C 242 for the ASTM definition of this
term. Most equipment manufacturers refer to this as the equivalent
spherical diameter.
D25
18h·log X1
X2
~ps–pf!u2t3108(1)
where:
h= viscosity of the fluid, cP,
X
1
= distance from the middle of the centrifugal disc cell to the point
in the cell where sedimentation starts,
X
2
= distance from the center of the centrifugal disc cell to the point
where the light beam intersects the particle
t= time for the particle to settle, s
r
s
= particle density, g/cm
3
,
r
f
= fluid density, g/cm
3
, and
u= rotational angular velocity, rad/s.
4.2 The instruments that have been found suitable for this
test method incorporate microcomputers that control instru-
ment operation and perform all required data acquisition and
computation functions.
5. Significance and Use
5.1 Manufacturers and users of advanced ceramic powders
will find this test method useful for determining the particle
size distribution of these materials for product specification,
quality control, and research and development.
6. Apparatus
6.1 Centrifugal Particle Size Distribution Analyzer—The
analyzer shall incorporate a centrifuge capable of subjecting an
homogeneous dispersion of the sample to centrifugal accelera-
tion in specially designed sample cells. A collimated beam of
1
This test method is under the jurisdiction of ASTM Committee C28 on
Advanced Ceramics and is the direct responsibility of Subcommittee C28.03 on
Physical Properties and Performance.
Current edition approved Jan. 1, 2008. Published January 2008. Originally
approved in 1994. Last previous edition approved in 2006 as C 1282 – 00 (2006).
2
For referenced ASTM standards, visit the ASTM website, www.astm.org, or
contact ASTM Customer Service at service@astm.org. For Annual Book of ASTM
Standards volume information, refer to the standard’s Document Summary page on
the ASTM website.
1
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
Copyright by ASTM Int'l (all rights reserved); Thu Apr 16 08:44:25 EDT 2009
Downloaded/printed by
Laurentian University pursuant to License Agreement. No further reproductions authorized.
visible light (either monochromatic or broad-band) shall
traverse the sample cell at a defined distance from the top of the
cell. The change in photo-extinction resulting from sedimen-
tation of the sample shall be measured by a photo-detector and
appropriate electronic circuits and used to calculate the
volume-based size distribution of the sample.
6.2 Ultrasonic Probe, consisting of a 200 to 300-W power
unit, ultrasonic transducer, and 13-mm (
1
⁄
2
-in.) diameter probe.
6.3 Balance, top-loading, accurate to 10 60.1 g.
6.4 Stirrer, magnetic, with 25-mm (1-in.) and 19-mm (
3
⁄
4
in.) stirring bars.
6.5 Temperature measurement device (thermometer, ther-
mocouple, etc.), 0 to 50°C, accurate to 0.5°C.
6.6 Sample Cells, as supplied by the instrument manufac-
turer.
7. Reagents
7.1 Purity of Reagents—Reagent grade chemicals shall be
used in all tests. Unless otherwise indicated, it is intended that
all reagents shall conform to the specifications of the Commit-
tee on Analytical Reagents of the American Chemical Society,
where such specifications are available.
3
7.2 Darvan C
4
Dispersing Fluid—Dissolve 5.0 g Darvan C
(ammonium salt of polymethacrylic acid) in 95 cm
3
distilled
water (Note 3).
NOTE 3—These reagents are adequate to cover the range of samples up
to a maximum particle diameter of 20 µm.
7.3 pH Adjusters—Fresh ammonium hydroxide (NH4OH)
at 5 to 10 % strength is a common reagent used to raise pH
while fresh nitric acid (HNO3) at 5 to 10 % strength is a
common reagent to lower pH.
8. Procedure
8.1 Sample Preparation and Dispersion:
8.1.1 Add the appropriate sample mass of powder, usually 1
g, to 250 cm
3
dispersing medium in a 500-cm
3
beaker. The
dispersing medium consists of 250 cm
3
of distilled water to
which 0.3 g of dispersing liquid is added.
8.1.1.1 Sample Dispersion—Place the beaker containing
sample in an ice water bath so that the temperature of the
sample does not exceed approximately 30°C. Adjust the pH of
the sample to 9.060.2 to ensure adequate dispersion aided by
the chemical environment. Deagglomerate the sample by
inserting the ultrasonic transducer into the dispersing liquid/
sample in a beaker. Ensure that the transducer is approximately
10.0 mm above the bottom of the sample beaker. Apply the
ultrasonic energy at 40 W or higher for 1.0 min, and allow the
sample to settle for 1.0 min. Repeat this procedure three times
so that the total ultrasonic energy application time is 3 min.
This high level of mechanical energy application is necessary
to accomplish deagglomeration of the particles. The extent of
deagglomeration is expected to be different for powders
produced by different processes. An alternate approach is to
ultrasonicate for 3 min continuously before pH and tempera-
ture check. Readjust the pH, if necessary, after ultrasonic
application. Keep stirring the sample continuously, and test it
as soon as practical to prevent reagglomeration.
8.1.2 Add a 25-mm (1-in.) stirring bar to the beaker, and
place it on a magnetic stirrer. Stir for approximately 3 min in
a cold water bath to bring the sample to ambient temperature.
Continue stirring at constant temperature.
8.2 Analyzer Preparation:
8.2.1 To warm up the analyzer, apply power for a minimum
of 10 min prior to testing. Conduct the warmup with the sample
compartment closed. Make certain that the ventilation airflow
is not restricted by adjacent equipment, papers, or other
materials. Check the printer to ensure a sufficient supply of
paper exists. Clean a pair of sample cells and caps, rinse with
the 0.1 % Darvan C solution, and store inverted on absorbent
paper.
NOTE 4—The concentration of the sample may require dilution with
0.1 % Darvan C solution to meet the optical absorbance tolerance
specified in the instrument operating manual. Thorough mixing must
accompany any dilution of the sample.
8.2.2 If required by the manufacturer’s operating manual,
check and adjust the zero and full-scale settings. These
adjustments can be crucial to reliable and reproducible mea-
surements.
8.2.3 Input the test parameters (sample and fluid density,
fluid viscosity, test range, sedimentation distance, and centri-
fuge speed) according to the manufacturer’s operating manual.
Set the fluid viscosity at the value corresponding to the actual
room temperature (see Table 1). Set the range such that the
volume percent greater than the maximum diameter is certainly
zero and the volume percent finer than the minimum diameter
is less than 10 % (cumulative percent oversize greater than
90 %). Preliminary test runs may be required to establish the
range and the size of the divisions necessary to achieve the test
limits. The test range and division settings shall provide a
minimum of ten divisions within the test range.
8.2.4 If required by the manufacturer’s operating manual,
conduct a blank test with clean dispersant fluid in the sample
and reference cells.
8.3 Test Performance:
8.3.1 Adjust the rate of the magnetic stirrer to produce a
slight vortex in the sample dispersion. Withdraw an appropriate
volume of sample from the center of the dilute dispersion with
a disposable plastic pipet. Make certain that the dispersion is
mixed thoroughly by vigorous pumping with the pipet (avoid
strong agitation, which would create bubbles). Completely
transfer the withdrawn sample to the drained sample cell. Insert
3
Reagent Chemicals, American Chemical Society Specifications , American
Chemical Society, Washington, DC. For suggestions on the testing of reagents not
listed by the American Chemical Society, see Reagent Chemicals and Standards,by
Joseph Rosin, D. Van Nostrand Co., Inc., New York, NY, and the United States
Pharmacopeia.
4
Darvan C is a trademarked product of R.T. Vanderbilt Company, Inc., Norwalk,
CT 06856-5150.
TABLE 1 Viscosity Coefficients of Water
A
Temperature, °C 20 21 22 23 24 25 26 27 28 29
Viscosity
coefficient,
mPa*s or cP
1.01 0.98 0.96 0.94 0.92 0.89 0.87 0.86 0.84 0.82
A
Viscosity coefficient values to two decimal places have been found satisfactory
for this test method.
C 1282 – 08
2
Copyright by ASTM Int'l (all rights reserved); Thu Apr 16 08:44:25 EDT 2009
Downloaded/printed by
Laurentian University pursuant to License Agreement. No further reproductions authorized.
the cell cap, carefully wipe all outer surfaces of the cell, place
in the centrifuge, and start the test. Monitor the absorbance
display on the instrument to ensure that the initial absorbance
is within the specified tolerance at the beginning of the test. If
not, refer to Note 4. When the test is completed, remove the
sample cell, clean immediately, and rinse with dispersant fluid.
9. Presentation of Data
9.1 Analyzer Printout—The analyzers that have been found
suitable for this test method render data printouts that provide
a permanent record of the test conditions and graphic records
of the size distribution. These printouts may provide sufficient
information for many purposes.
10. Fluid Densities and Viscosity Coefficients
10.1 Table 1 lists the viscosity coefficients for water over the
temperature range of 20 to 29°C. Use these values for the
Darvan C solution. Use a density of 1.00 g/cm
3
for these
solutions.
11. Additional Data Presentations
11.1 The analyzers that have been found suitable for use
with this test method produce a printed report that may include
a cumulative size distribution table showing the volume
percent oversize, starting with the programmed maximum
diameter. The incremental divisions are selectable by the
operator in some operating modes and, in others, determined
by the microcomputer in the instrument.
12. Precision and Bias
12.1 Precision—No statement of precision can be made
since round-robin testing has not been completed.
12.2 Bias—It is not possible to discuss the bias of the results
obtained using this test method since no absolute method of
particle size distribution is recognized.
13. Keywords
13.1 advanced ceramics; particle size distribution;
photosedimentation
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).
C 1282 – 08
3
Copyright by ASTM Int'l (all rights reserved); Thu Apr 16 08:44:25 EDT 2009
Downloaded/printed by
Laurentian University pursuant to License Agreement. No further reproductions authorized.
摘要:

ASTM C1282 - 08 是美国材料与试验协会(ASTM)发布的一项标准规范,全称为“Standard Specification for Polyolefin and Polyethylene (PE) and Polyvinyl Chloride (PVC) Sealing Tapes for Use in Conjunction with Insulation and Vapor Retarders”。该标准主要规定了用于建筑隔热层与防潮层配合使用的聚烯烃、聚乙烯及聚氯乙烯密封胶带的

展开>> 收起<<
ASTM C1282 - 08.pdf

共3页,预览3页

还剩页未读, 继续阅读

声明:本文档系会员上传,若文档所含内容侵犯了您的版权或隐私,请立即通知,我们立即给予侵权申诉删除!
作者:Carl 分类:国外协会 价格:15星币 属性:3 页 大小:64.36KB 格式:PDF 时间:2024-09-02

开通VIP享超值会员特权

  • 多端同步记录
  • 高速下载文档
  • 免费文档工具
  • 分享文档赚钱
  • 每日登录抽奖
  • 优质衍生服务
/ 3
客服
关注