an exciting coil and sensing coil or Hall element mounted in a
common holder. A Hall element is a semiconductor that by
reason of the Hall effect is capable of responding in a manner
directly proportional to magnetic-flux density. However, when
used with an exciting coil, it should be remembered that eddy
current flow is influenced by the excitation frequency.
NOTE 3—Lift-off effect is an important consideration since coupling
variations can affect the examination significantly.
7.4 Driving Mechanism—A mechanical device capable of
passing the tube through the examination coil or past the probe.
It shall operate at a uniform speed with minimum vibration of
coil, probe, or tube and maintain the article being examined in
proper register or concentricity with the probe or examination
coil. Where required, the mechanism shall be capable of
uniformly rotating the tube or probe.
7.5 Reference Standard—The standard used to adjust the
sensitivity setting of the apparatus shall be sound and of the
same nominal alloy, temper, and nominal dimensions as the lot
of tubes or pipes to be examined on a production basis. It shall
be of sufficient length to permit the required spacing of the
artificial discontinuities (at least 4 ft, and preferably longer).
Artificial discontinuities made in the tube or pipe shall be
centered as nearly as possible on one inside or outside diameter
surface of the tube and shall preferably be of one of the
following types:
7.5.1 Holes—Holes that are usually drilled completely
through the wall may be used. Care should be taken during
drilling to avoid distortion of the tube (or pipe) and hole.
7.5.2 Notches—Notches may be produced by electric dis-
charge machining (EDM), milling, or other means. Longitudi-
nal or transverse notches or both may be used (Note 4).
Orientation, dimensions (width, length, and depth), and con-
figuration of the notches affect the response of the eddy current
system. Notch depth is usually specified as a percentage of
nominal wall thickness of the tubular product being examined.
Notches may be placed on the outer, inner, or both surfaces of
the reference (calibration) standard. Outer surface notches
provide an indication of system response to discontinuities
originating on the outer tube surface, whereas inner surface
notches provide an indication of system response to disconti-
nuities originating on the inner tube surface (Note 5).
NOTE 4—Longitudinal notch standards are normally used when exam-
ining with rotating probe systems.
NOTE 5—The density of eddy currents decreases nearly exponentially
with increasing distance from the surface nearest the coil, and the
sensitivity to subsurface discontinuities decreases with the change in depth
allowing the use of phase analysis techniques.
7.5.3 The configuration, orientation, and dimensions (diam-
eter of holes and the width, length, and depth of notches) of the
artificial discontinuities to be used for establishing acceptance
limits should be subject to agreement between supplier and
purchaser.
8. Adjustment and Standardization of Apparatus
Sensitivity
8.1 Select the apparatus, examination frequency, coil or
probe, or both, design, phase discrimination, and other circuitry
as well as speed of examining which shall demonstrate the
system capability for detecting the discontinuities of interest.
8.2 Fabricate the applicable reference standard in accor-
dance with the agreement between the purchaser and tubing
supplier.
8.3 Adjust the apparatus to obtain an optimum signal-to-
noise ratio with the minimum sensitivity required to detect the
artificial discontinuities in the reference standard. Do this
under conditions (such as examining speed) identical to those
to be used in production examination of the tubular products.
8.4 Determine the end effect by using a special reference
tube or pipe containing a series of notches or holes near one or
both of the ends and passing this reference standard through the
system at production examination speeds. If notches or holes
are placed near only one of the ends, pass the tube through the
system backwards and forwards.
9. Procedure
9.1 Standardize the apparatus at the start of the examination
run using the reference standard. The recommended maximum
interval between re-standardization is four hours although
more or less frequent re-standardization may be done by
agreement between using parties, or whenever improper func-
tioning of the equipment is suspected. If improper functioning
is found, re-standardize the apparatus and re-examine all tubes
or pipes examined during the period since the last successful
standardization.
9.2 Pass the lot of tubes or pipes to be examined through the
examination coil or past the probe coil of the apparatus
adjusted to the sensitivity as described in Section 8. Set aside
tubes or pipes with discontinuities indicated by the apparatus.
It is recommended that tubes with discontinuity indications be
re-examined in accordance with the purchase specification.
9.3 Tubes or pipes may be examined in the final drawn,
annealed, heat treated, as-welded, or other step in processing.
The point in processing at which examination is made should
be agreed upon by the supplier and the purchaser. The tubes
should be free of any substance that may interfere with the
examination.
10. Supplemental Information Regarding Eddy Current
Examination of High Alloy Steels and Similar Alloys
10.1 In the eddy current examination of austenitic
chromium-nickel stainless steels it has been found that exami-
nation frequencies ranging from less than 1 kHz to more than
1 MHz can be used. The more commonly used operating
frequencies are in the range, 1 kHz to 125 kHz. The exact
frequency used will depend on the application.
Example—If thin-walled tubular products are to be exam-
ined or surface rather than subsurface discontinuities are to be
detected, higher frequencies are used. For heavy-walled tubes,
subsurface discontinuities well below the surface require the
use of a lower frequency. Choice of examination frequency
will determine the size of discontinuity that can be detected.
10.2 As in any eddy current examination, the depth and
orientation of the discontinuity below the entry surface will
E426 − 16 (2021)
3