3.3.2 cumulative failure probability, p
f
—the probability that
a single selected specimen chosen at random from a population
of specimens will fail at or before reaching the K
Jc
value of
interest.
3.3.3 extent of slow-stable cracking prior to cleavage
initiation, ∆a
sc
[L]—the average amount of any ductile crack
growth that precedes rapid cleavage failure or test termination.
3.3.4 load-line displacement rate, ∆
˙
LL
[LT
-1
]—rate of in-
crease of specimen load-line displacement.
3.3.5 provisional reference temperature, T
0Q
[°C]—
provisional T
0
value calculated using the standard test method
described herein.
3.3.5.1 Discussion—T
0Q
is validated as T
0
in 10.5.
3.3.6 reference temperature, T
0
[°C]—The test temperature
at which the median of the K
Jc
distribution from 1T size
specimens equals 100 MPa√m (91.0 ksi√in.).
3.3.7 specimen size, nT—a code used to define specimen
dimensions, where nis expressed in multiples of 1 in.
3.3.7.1 Discussion—In this method, specimen proportional-
ity is required. For compact specimens and bend bars, speci-
men thickness B = n inches. The metric equivalent of ncan
either be defined as 25.4 mm or rounded down to 25 mm.
3.3.8 time to control force, t
m
[T]—time to P
m
.
3.3.9 Weibull fitting parameter, K
0
—a scale parameter cor-
responding to the 63.2 % cumulative failure probability level
(4).
3.3.9.1 Discussion—K
Jc
= K
0
when p
f
= 0.632.
3.3.10 Weibull slope, b—slope of the p
f
versus K
Jc
fitting
line in a Weibull plot where data pairs are plotted in linearized
coordinates
3.3.10.1 Discussion—values of p
f
are calculated by means
of Eq 22.
3.3.10.2 Discussion—A fitting line with slope bdefines the
characteristic of the typical scatter of K
Jc
data.
3.3.10.3 Discussion—A Weibull slope of 4 is used exclu-
sively in this test method.
4. Summary of Test Method
4.1 This is a general overview of the major aspects associ-
ated with determining T
0
for a ferritic steel. It is not intended
to provide a prescriptive method to be rigorously followed
sequentially. The general steps necessary for conducting a test
are provided within the following sub-sections. A graphical
overview of the test method is presented in Fig. 1 in the form
of a flowchart.
5
4.2 Select the configuration and dimensions of the test
specimens, 7.1-7.3. This test method covers C(T), DC(T), and
SE(B) specimens. The specimen dimension requirement, based
on the maximum K
Jc
capacity of a specimen K
Jclimit
,Eq 3, is
provided in 7.5, and depends on the values of σ
YS
and Eof the
steel at the test temperature.
4.3 Select the test temperature(s) for the initial test(s) within
a temperature range where K
Jc(med)
for the specimen size tested
is approximately 100 MPa√m, 8.4. If available, Charpy
V-notch data can be used to establish the initial test
temperature(s), 8.4.1 and Table 3. When using small specimens
(precracked Charpy or smaller), it is generally advisable to test
below the expected T
0
,8.5.
4.4 Establish the number of test specimens, based on the
available material and the requirements for obtaining a valid
T
0
,10.3. A minimum of six tests are required by this test
method, (5)and results are weighted based on the difference
between test temperatures and T
0
, Table 4. An iterative
procedure is suggested in 10.3.1 for identifying an appropriate
sequence of test temperatures.
4.5 Precrack specimens by fatigue, 7.8. Methods to promote
early crack initiation are provided in 7.8.1, which also includes
maximum values for the control force, P
m
,Eq 5 or Eq 6. The
fatigue precracking procedure is provided in 7.8.2. Specimens
may be optionally side-grooved in accordance with 7.7, pref-
erably after precracking, in order to limit the curvature of the
initial crack front and avoid tunneling, hence fulfilling the
crack curvature requirements of 8.9.1.
4.6 Perform initial test(s) at the temperature(s) selected in
4.3. Tests shall be performed in accordance with the procedure
outlined in Section 8, and the corresponding provisions given
in Test Method E1820 (8.5 for Optical Crack Size
Measurements, 8.3.1 for Bend Testing, and 8.3.2 for Compact
Testing). Provisions for temperature control and measurement
are given in 8.6. Requirements for quasi-static loading rate are
provided in 8.7.1. For loading rates outside the quasi-static
range, refer to 8.7.2 (slow loading rates) or Annex A1 (elevated
loading rates).
4.7 For each test performed, calculate the J-integral at the
onset of cleavage fracture, J
c
, for the relevant specimen type as
prescribed in 9.1.1-9.1.4. Determine the corresponding value
K
Jc
according to 9.1.5. If a significant pop-in is observed or
suspected, establish its significant in accordance with 8.9.4 and
9.2.
4.8 If the K
Jc
value exceeds K
Jclimit
in Eq 3 and/or the
ductile crack extension limit, K
Jc∆a
(see 8.9.2) is violated,
perform data censoring according to 8.9.2 and 10.2.2. If
cleavage fracture does not occur, refer to 8.9.3.
4.9 Size correct each K
Jc
value to 1T equivalence using the
measured specimen thickness in Eq 22,10.2.3.
4.10 Determine the provisional reference temperature, T
0Q
,
following the iterative procedure in 10.2.4 (tests performed at
multiple temperatures) or 10.4 (tests performed at a single
temperature), and accounting for the T
0Q
650 °C test
temperature requirement in 10.2.4.
4.11 If necessary, adjust test temperature(s) and perform
additional tests (if possible, given the number of remaining
specimens). For each additional test performed, calculate J
c
and K
Jc
according to Section 9. If needed, censor (8.9.2 and
5
The flowchart illustrated in Fig. 1 is only meant to capture the principal
components of an analysis conducted in accordance with this test method, and not
to address all possible outcomes and requirements. Other pathways through the
standard are possible and appropriate, given the constraints and objectives of a
specific analysis.