Designation: E319 −85 (Reapproved 2014)
Standard Practice for the
Evaluation of Single-Pan Mechanical Balances
1
This standard is issued under the fixed designation E319; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (´) indicates an editorial change since the last revision or reapproval.
INTRODUCTION
The balance performs two basic functions: (1) it compares an unknown load with one or more
weights, and (2) it indicates the difference between the two loads for differences smaller than the
smallest weights normally used on the balance. The test procedure given herein measures the precision
with which the balance can compare the two loads, and the rates at which systematic errors may affect
the observed difference.
1. Scope
1.1 This practice covers testing procedures for evaluating
the performance of single-arm balances required by ASTM
standards.
1.2 This practice is intended for but not limited to sensitivity
ratios of 10
6
or better and on-scale ranges of 1000xd or more
where d= reability either directly or by estimation.
1.3 This practice can also be applied to other single-pan
balances with mechanical weight changing of different capaci-
ties or sensitivities with appropriate test loads and calibration
weights.
NOTE 1—Mechanical balances of this type have largely been replaced
by automatic electronic balances incorporating a variety of operational
principles. Nevertheless, some single-pan mechanical balances are still
manufactured and many older balances will remain in service for years to
come. One type of automatic electronic balance, the so-called “hybrid,”
bears considerable similarity to single-pan mechanical balances of the null
type. (1)
2
1.4 This standard does not purport to address all of the
safety problems, if any, associated with its use. It is the
responsibility of the user of this standard to establish appro-
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use.
2. Terminology
2.1 Definitions(1):
2.1.1 accuracy—the degree of agreement of the measure-
ments with the true value of the magnitude of the quantity
measured (2).
2.1.2 correction for a weight—the correction for the error in
adjustment is:
Cr.W 5A2N(1)
where:
Cr.W = correction for the error in adjustment to nominal
value,
A= actual value of the weight, and
N= nominal value.
NOTE 2—In practice it is not possible to adjust weights exactly to their
nominal values.
2.1.3 correction for error in scale indication— the correc-
tion for the scale indication, I, is:
Cr.I 5A2I(2)
NOTE 3—The correction for the scale is taken with reference to the
measured value of a weight used as a test load during calibration of the
on-scale range.
2.1.4 index of precision—the standard deviation, computed
in any acceptable manner, for a collection of measurements
involving a given pair of mass standards (3).
NOTE 4—The standard deviation is computed from the data provided by
the instrument precision test (see Section 7) index of precision.
2.1.5 null-type balance—a balance which requires, as the
final step in its operation, that the observer restore the angle of
the balance beam to its original (or null) position. The least
significant figures of the balance indication are obtained from
this operation.
2.1.6 optical-type balance—in this type the least significant
figures of the balance indication are related to the deflection
angle of the beam from its original (or null) position. A scale
placed on the moving beam is optically projected onto the
(stationary) balance case to provide this indication.
1
This practice is under the jurisdiction of ASTM Committee E41 on Laboratory
Apparatus and is the direct responsibility of Subcommittee E41.06 on Laboratory
Instruments and Equipment.
Current edition approved Nov. 1, 2014. Published November 2014. Originally
approved in 1968. Last previous edition approved in 2008 as E319 – 85 (2008).
DOI: 10.1520/E0319-85R14.
2
The boldface numbers in parentheses refer to the list of references at the end of
this practice.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
2.1.7 precision—the repeatability of the balance indication
with the same load under essentially the same conditions.
NOTE 5—The more closely the measurements are grouped, the smaller
the index of precision will be. The precision must be measured under
environmental conditions that represent the conditions under which the
balance is normally used.
2.1.8 readability—the value of the smallest decimal subdi-
vision of a scale division in terms of mass units, that can be
read, when the balance is read in the intended manner.
NOTE 6—The readability of a particular instrument is not a measure of
its performance as a weighing device. The relationship between the
numerical value obtained by reading devices and the ability of the operator
to estimate the location of the reference device or index is important. It is
possible to introduce a large number of readable subdivisions of the main
scale divisions that would increase the “readability” as defined but if the
reading device cannot be reset to the same numerical value when the beam
is in an immovable condition, or when the load on the beam is a constant
value, the readability becomes meaningless. Readability substantially less
than 1 standard deviation as determined by repeated measurement with a
given test weight is usually superfluous.
2.1.9 scale division—the smallest graduated interval subdi-
vided either by estimation or with the aid of a vernier.
Subdivisions which appear as divisions on the vernier are not
considered to be scale divisions, but rather parts of scale
divisions.
2.1.10 sensitivity weight—a small weight used to measure
the “on-scale” deflection of the balance indicator.
NOTE 7—With single-pan balances the sensitivity weight should be
equal to the value of the smallest built-in weight represented by the first
step on the dial for the smallest weights.
2.1.11 test load—a load chosen to represent the sample load
in the test procedure.
2.1.12 value of the division—the change in load required to
change the balance indication by one scale division. The
reciprocal of the sensitivity is its most useful function.
3. Summary of Practice
3.1 The accuracy of the direct-reading scale, the smallest
weight of the set of built-in weights, and uniformity of
sensitivity between the upper and lower halves of full-scale
deflections are verified by preliminary tests.
3.2 Estimates of rate of change of the zero with time, rate of
change in the value of the scale division with time, and a
quantitative measure of the variability or random error are
provided by short tests for precision and bias.
3.3 An overall test of the direct-reading capability is pro-
vided by tests of the built-in weights.
4. Significance and Use
4.1 Monitoring Weighing Performance—This practice pro-
vides results in the form of control charts which measure the
weighing capability at the time of the test. A series of tests at
appropriate intervals will monitor balance performance over a
period of time. A marked change from expected performance
may result from a variety of causes including: maladjustment,
damage, dirt, foreign material, and thermal disturbances. If the
test results are to indicate future performance, any disturbances
that occur exterior to the balance must be brought under control
(2).
4.2 Acceptance Tests—This practice may also be used as
acceptance tests for new balances. For this purpose, the tests
should be conducted under favorable, but not necessarily ideal,
conditions. Since systematic error in the course of the zero and
the course of the sensitivity may be caused by disturbances
external to the balance, limits on these errors are not ordinarily
prescribed in acceptance requirements.
5. Preparation of Apparatus
5.1 Balance (In all cases, the balance should be used in
accordance with the manufacturer’s instructions):
5.1.1 The results obtained will depend on the environment.
Select an area which is free of excessive vibration and air
currents, where rapid changes in temperature and relative
humidity will not be encountered, and where the floor is rigid
enough to be free of a tilting effect on the balance indication.
Place the balance on a sturdy bench. If the balance has been
moved to a new location, permit it to come to thermal
equilibrium for at least 1 h before performing the test,
preferably several hours.
5.1.2 Inspect and test the balance to make sure that it is in
proper mechanical order. Arrest and release the beam to make
sure that readings are approximately repetitive. Observe the
indication during arrest and release to ensure that there is no
“kick” that would indicate that arrestment points might be out
of adjustment. If necessary, have the balance adjusted by a
competent balance technician.
5.1.3 Make a few trial measurements of the interval from
zero to the full-scale indication.
5.2 Reading the Balance—The balance should be read in
accordance with the instructions supplied by the manufacturer.
Optical types should include the reading of verniers or mi-
crometres. Null types should include the indication of the
device for restoring to null including verniers or micrometres.
6. Preliminary Testing of Single-Arm Balances
6.1 Summary of Method—With single-pan balances the
smallest built-in weight, indicated by the first step on the dial,
is compared with a calibrated weight. The direct-reading scale
is tested for agreement with the smallest built-in weight and the
sensitivity is adjusted, if necessary, so that the indications of
the scale are precise in terms of the calibrated weight. A
“fifty-fifty” test verifies the accuracy of the midpoint at
half-full scale. This test should be performed before proceed-
ing to other tests. After the accuracy of adjustment of the
smallest built-in weight is verified, this weight is used to test
full-scale deflections. Tests are also made for the uniformity of
deflection over the lower and upper halves of the full-scale
deflection. The preliminary tests show either that the balance is
operating properly, or that discrepancies indicate the presence
of sources of error. Uncertainties of perhaps one millionth of
the balance capacity may be caused by dirt or foreign material
in the bearings, or by unskilled handling, while larger discrep-
ancies may be caused by worn or damaged knife-edges or other
sources such as electrostatic effects. Any necessary cleaning or
E319 − 85 (2014)
2
servicing should be done at this point. If discrepancies
continue, other possible sources of uncertainty should be
studied. There is no point in proceeding with routine test
procedures until acceptable results can be obtained with the
preliminary tests.
NOTE 8—With null-type balances (including the hybrid) it is possible to
use the flexure of a segment of metal, quartz, etc. as the main pivots
instead of knife edges. A flexure pivot is by its nature free of problems of
dirt. Flexures are also generally more robust than knives. The chief
problem associated with flexures is that they act like springs and thus add
a restoring force which may vary with time or temperature. This drawback
can be minimized by careful design and all but eliminated by the use of
servo-control in electronic balances.
6.2 Materials:
6.2.1 A watch or clock which indicates seconds,
6.2.2 Pencils for recording data,
6.2.3 Columnar data sheets (If balance performance will be
monitored periodically, it may be useful to enter data directly
into a personal computer which has been programmed for this
task.),
6.2.4 A calibrated weight designated S1 which has the
nominal value equal to the smallest interval on the dial-
operated weights, and
6.2.5 Two weights of half of the nominal value of S1
designated (
1
⁄
2
)1 and (
1
⁄
2
)2. (These weights need not be
calibrated but they should bear distinguishing marks, prefer-
ably one, and two dots.)
6.3 Procedure—Adjust the “no-load” readings to a point
near the center of the vernier so that zero drift or other
deviation will not cause a negative scale reading. Perform the
preliminary tests, loading the pan and changing the dial
settings according to the schedule in Table 1. Before releasing
the beam, record the load on the pan and the dial setting so that
the observation will be confined to the scale reading. Release
the balance and observe the scale reading. Record the indica-
tion and verify the stability of the scale reading, then arrest the
balance promptly.
6.4 Calculations for Preliminary Tests:
6.4.1 Compute D1, the value of the smallest built-in weight
as follows:
D15
@
~
a2b1f2e
!
/2
#
1S1 (3)
where: a, b, f, and eare taken from Table 1, and S1 = cali-
brated value of test weight.
6.4.2 Compute S1 in scale divisions to verify the full-scale
value on the direct-reading scale as follows:
S15
~
c2b1d2e
!
/2 (4)
where c, b, d, and eare taken from Table 1. Adjust the
balance sensitivity if necessary so that the full-scale reading
equals D1.
6.4.3 Compute average scale difference, A, for lower 50 %
of direct-reading scale as follows:
A5
~
g2f1j2k
!
/2 (5)
6.4.4 Compute average scale difference, B, for upper 50 %
of direct-reading scale as follows:
B5
~
h2g1i2j
!
/2 (6)
A and Bshould agree within 3 standard deviations (see
7.5.3). Any discrepancy smaller than 3 standard deviations may
be ascribed to uncertainty in the preliminary measurements and
does not necessarily indicate a real change in the value of the
scale divisions.
6.4.5 Inspect the no-load readings, a, f, and kfor agreement
or zero drift.
6.4.6 See Table 2 and Fig. 1 for examples of calculations
and observation form.
7. Instrument Precision (4)
7.1 Summary of Method:
7.1.1 A set of four readings is repeated four times, or more,
to obtain pairs of readings with identical loads:
7.1.1.1 A reading near zero,
7.1.1.2 A reading near the upper end of the scale,
7.1.1.3 A reading near the upper end of the scale with a test
load plus a small weight, and
7.1.1.4 A reading near zero with the test load but with the
small weight removed.
7.1.2 Readings are taken at a steady pace as rapidly as
practicable, consistent with good practice, and the time is
observed at the start of each set of observations and at the end
of the test.
7.1.3 The balance indications are plotted on a graph to
provide a visual presentation of errors. The zero readings are
connected to show the course of the zero with time. The
response of the balance to the small weight is plotted. The
course of the sensitivity with time is represented by a plot of
TABLE 1 Schedule for Preliminary Tests of Single-Arm Balances
Observation Time Pan Load Dial Setting Scale Reading
a Record the time zero 0 ...
bS1
A
1 ...
cS1
A
0 ...
dWait30sS1
A
0 ...
eS1
A
1 ...
f Record the time 0 0 ...
g(
1
⁄
2
)
1B
0 ...
h Add (
1
⁄
2
)
2B
(
1
⁄
2
)
1
+(
1
⁄
2
)
2B
0 ...
iWait30s(
1
⁄
2
)
1
+(
1
⁄
2
)
2B
0 ...
j Remove (
1
⁄
2
)
1B
(
1
⁄
2
)
2B
0 ...
k Record the time 0 0 ...
A
S1 = calibrated weight of nominal value equal to the smallest dial-operated weight.
B
(
1
⁄
2
)
1
and (
1
⁄
2
)
2
= weights of nominal value equal to
1
⁄
2
S1 (not necessarily calibrated but marked for identification).
E319 − 85 (2014)
3
摘要:

ASTM E319 - 85 (2014) 是一项由美国材料与试验协会(ASTM)制定的标准规范,标题为“Standard Practice for Fracture Testing of Ferritic-Bainitic Steels with Low Fracture Toughness”(测定低断裂韧性铁素体-贝氏体钢的断裂试验标准规程),该标准首次发布于1985年,并于2014年经过重新批准确认。它详细规定了在低温或高应变速率条件下,对具有较低断裂韧性水平的铁素体-贝氏体钢进行断裂力学

展开>> 收起<<
ASTM E319 - 85 (2014).pdf

共9页,预览4页

还剩页未读, 继续阅读

声明:本文档系会员上传,若文档所含内容侵犯了您的版权或隐私,请立即通知,我们立即给予侵权申诉删除!
作者:Carl 分类:国外协会 价格:10星币 属性:9 页 大小:300.69KB 格式:PDF 时间:2024-09-04

开通VIP享超值会员特权

  • 多端同步记录
  • 高速下载文档
  • 免费文档工具
  • 分享文档赚钱
  • 每日登录抽奖
  • 优质衍生服务
/ 9
客服
关注