3.2.2 refractory metal, n—a metal, such as hafnium,
molybdenum, niobium, tantalum, vanadium, or tungsten, char-
acterized by very high melting points, above about 1900 °C,
that oxidizes at temperatures far below its melting point.
4. Significance and Use
4.1 Test methods for chemical analysis of reactive and
refractory metals are primarily intended to test such materials
for compliance with compositional specifications. It is assumed
that all who use this guide will be trained analysts capable of
performing common laboratory procedures skillfully and
safely. It is expected that work will be performed in a properly
equipped laboratory under appropriate quality control practices
such as those described in Guide E882 and Practice E2437,as
well as in ISO 17025 and ISO Guide 32.
4.2 This guide is intended to aid analytical chemistry
laboratories in the analysis of reactive and refractory metals
and their alloys when no specific standard test methods are
available. The principles incorporated in this practice can also
be applied in laboratories that wish to validate and document
non-standard test methods.
4.3 The analysis of reactive and refractory materials is
typically performed by only a small number of laboratories.
Few of these laboratories have analytical instrumentation in
common for use in interlaboratory proficiency testing pro-
grams. This requires the use of within-laboratory developed
test methods that vary between laboratories. It is intended that
this practice will give general guidance to experienced person-
nel that will assist them in the development of a procedure that
will meet their analytical objectives.
4.4 Practice E2438 provides guidance for the development
and documentation of an In-House Standard Operating Proce-
dure (SOP).
5. Hazards
5.1 The dissolution of these metals usually requires the use
of hydrofluoric acid. Read and follow label precautions and
mandatory safety data sheet (MSDS) information, and refer to
Practices E50.
5.2 Fine turnings, chips, or powder require more water to
moderate the reaction rate. If the metal is solid chunks rather
than finely divided, the dissolution will be much slower and
require less water. It will also require additional acid to
complete the dissolution. See Specification D1193 for reagent
water specifications.
5.3 Fine turnings or powder of hafnium, titanium, and
zirconium are pyrophoric. Observe the proper precautions.
6. Test Methods
6.1 Atomic Absorption Spectrometry (AAS)—This technique
is applicable to elements that can be dissolved and placed in the
vapor state (flame or flameless) as ground state atoms. Radia-
tion from a hollow cathode light source emits the spectrum of
the element to be analyzed, which is then passed through vapor
containing the element to be analyzed, and into a monochro-
mator set for radiation characteristic of the element. The degree
of absorption is measured, and is proportional to the amount of
the particular element present. Analysis is made by comparison
to reference materials. Lists of elements that can be
determined, sensitivities, wavelengths, and operating param-
eters are provided by the instrument manufacturers. Due to the
requirements of the materials specifications, the AAS detection
limits for many elements will not be low enough for the use of
this technique. AAS has been utilized for the analysis of iron,
chromium, nickel, and tin in zirconium alloys and for the
determination of iron and zirconium in hafnium. Practices
E1184 and E1770 provide further details.
6.2 Atomic Emission Spectrometry (AES)—This technique
has historically been the main means of analysis for these
metals. Analysis by the techniques described in this section is
based upon the fact that each vaporized element, when suitably
excited by an arc, spark, plasma, or other means, is made to
emit its characteristic spectrum. The radiation thus emitted
passes into a spectrometer where it is resolved into its
component wavelengths and recorded by various means as a
definite pattern. The position of the spectral information
indicates QUALITATIVELY which elements are present. The
intensity determines QUANTITATIVELY the amount of each
element in the sample when compared to intensities obtained
by using calibration materials. Calibration materials can be
reference materials or industry acceptable analyzed specimens.
6.2.1 Arc/Spark Spectrometry (AS-AES)—Instruments uti-
lizing photographic film or plates were the mainstay for many
years until parts and supplies became very difficult, if not
impossible, to obtain. Electronic read-out arc/spark instruments
are still used by a number of laboratories. The primary problem
with arc/spark is the unavailability of solid reference materials
for calibration. Metallurgical history and surface grain size
effects can become an issue. However, if proper calibration
materials were available, the majority of the necessary ele-
ments in a typical material specification could be analyzed.
6.2.2 Direct Current Plasma Atomic Emission Spectrometry
(DCP-AES)—Since DCP-AES analysis, like atomic
absorption, Inductively Coupled Plasma Atomic Emission
Spectrometry (ICP-AES) (see 6.2.3), and Inductively Coupled
Plasma-Mass Spectrometry (ICP-MS) (see 6.2.4), relies on
sample dissolution, the major difficulties with reference mate-
rials for calibration with can be overcome by the techniques
discussed in 9.4. One of the main positive attributes of this
technique is the high resolution available due to the Echelle
grating. The highly complex spectra of reactive and refractory
metals require an instrument with good spectral resolving
power in order to find interference-free wavelengths. The
DCP-AES has a rugged sample introduction system that lends
itself well to the analysis of materials that have been dissolved
in acids such as hydrofluoric, nitric, and hydrochloric acids. In
most cases, due to the nature of the source design, the
DCP-AES does not achieve the detection limits or the precision
of an Inductively Coupled Plasma (ICP) source. In some cases,
however, its significantly superior resolving power may allow
it to out-perform an ICP that utilizes a conventional ruled
grating. The corrosion resistant nature of the sample introduc-
tion system and jet materials on DCP-AES has allowed it to be
very capable for the analysis of silicon in a hydrofluoric acid